Optimum for the auto measurement of Ultra-low Resistor
● Optimum for the Ultra-low resistance measurement such as shunt resistor.
● Measurement not to influenced by thermal electro motive force.
● Impressed the measuring current at pulse interval, in order to reduce abrasion of measuring terminal.
● Available to percentage measurement for the setup basic value:0.5mΩ~1kΩ/±50.00% [Resolution:50nΩ]
● GP-IB/RS-232C/Centoronics outputs are available (option)
● Comparison result by built-in comparator is open-collector output and displayed by LED and buzzer.
● Standard equipments of check circuit for measuring current and abnormal voltage
● Available to shift output
Specifications
Measuring range and Accuracy (at23℃±5℃)
Range | Measuring range | Resolution | Measuring current | Accuracy [Slow] |
---|---|---|---|---|
1mΩ | 0.0000mΩ~1.5000mΩ | 0.1μΩ | 3A | within ±(0.01%rdg+1μΩ) ±3digit [Average] ±4digit [Slow] ±5digit [Fast] |
10mΩ | 0.000mΩ~15.000mΩ | 1μΩ | 1A | |
100mΩ | 0.00mΩ~150.00mΩ | 10μΩ | ||
1Ω | 0.0000Ω~1.5000Ω | 100μΩ | 100mA | |
10Ω | 0.000Ω~15.000Ω | 1mΩ | ||
100Ω | 0.00Ω~150.00Ω | 10mΩ | 10mA | |
1kΩ | 0.0000kΩ~1.5000kΩ | 100mΩ | 1mA | |
% | 0.5mΩ~1kΩ/±50.0% | 0.01% [50nΩ] | Refer to the above |
Open-circuit voltage of measuring terminal | less than 5V |
---|---|
Measuring method | 4-terminal measuring method (contact check available) |
Sampling time | [Free running mode] 2~10 times/sec. |
[Remote start mode] about 16.5msec./29.8msec. | |
Setting range for comparator | Upper limit, lower limit for both [0~15000], % range:50.00% [low:0~-50.00%, high:0~+50.00%] |
Indication of comparator’s judgment result | LED indication LO/GO/HI and buzzer |
Control signal | Remote start input:”L” [0V] →”H” [DC12V] start |
Remote hold input:Open and “H” [DC12V]:Free run/”L” [0V]:Hold | |
comparison output [LO/GO/HI]:open collector:max.40V, 100mA | |
contact error output [CE]:open collector:max.40V, 100mA | |
end of comparison output [EOC]:open collector:max.40V, 100mA | |
Operation condition | [Temp.] +5℃~+40℃ [Humidity] less than 85% |
Power supply | AC100V/117V/220V/240V selectable, 50/60Hz, about 60VA |
Outer dimension | about 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.) |
Weight | about 3.8kg |
The Outline
AX-1152D can measure the ultra low resistance from 0.0000mΩ to 1.5000kΩ with high speed, high accuracy.
The unit has DOUBLE Measuring Mode can cancel to measure the thermo electro motive force that caused an error especially when measures ultra low resistance, and specail auto zero circuit, available to measure high accuracy, and high stability.
The unit is indicated the measured value as a digital and outputs signal outside to judge HI, GO, LO decision. It can be switched a measuring speed, selectable from AVERAGE/DOUBLE/SLOW/FAST.
And Contact check circuit and check circuit for measuring current and abnormal voltage are standard equipment, improved the reliability for the measurement furthermore.
The unit can be equipped either one GP-IB/RS-232C/centronics as an option.