Compare with external standard thermistor, percentage measurement
● Available to make % measurement against standard value at room temperature regardless of temperature characterristice in comarison with external standard-termistor.
● Possible to reduce error of the measured value for Self-heating on the measuring object by pulse impress of low measuring power (Within 150μW).
● Available to measure the resistance value of external standard or to measure the resistance value of the measuring object.
● Available to calibrate the difference (±9.999%) on resistance value of external standard and to correctly measure the measuring object.
● Digital comparator is built-in.
● Measuring current/voltage-check are built-in as standard function.
Specifications
Measuring range and Accuracy (at23℃±5℃)
Range | Measuring range | Measuring current | Accuracy | Display |
---|---|---|---|---|
10Ω | 0.000Ω~15.000Ω | 3mA | within ±0.03%±2digit [Slow]/within ±5digit [FAST] | 15000 |
100Ω | 0.00Ω~150.00Ω | 1mA | within ±0.02%±2digit [Slow]/within ±5digit [FAST] | |
1kΩ | 0.0Ω~1500.0kΩ | 300μA | ||
10kΩ | 0.000kΩ~15.000kΩ | 100μA | ||
100kΩ | 0.00kΩ~150.00kΩ | 10μA | ||
1MΩ | 0.0kΩ~1500.0kΩ | 1μA | within ±0.02%±3digit [Slow] | |
5MΩ | 0kΩ~5000kΩ | 1μA | within ±0.02%±5digit [Slow] | 5000 |
% | 1Ω~5000kΩ | refer to the above | refer to the above | 5000 |
±50.00% |
Open-circuit voltage of measuring terminal | less than 12V |
---|---|
Measuring method | 4-terminal measurement, Scanning of Rs/Rx, Available to Contact-Check both Rs and Rx. |
Sampling time | [Free running mode] 5 times/sec. [Slow], 10 times/sec. [FAST] [If the average mode] 1.8 times/sec. [Slow], 3.7 times/sec. [FAST] |
[Remote start mode] about 49msec. [Slow], about 24msec [Fast] (Difference by range and mode.) | |
Comparator set range | ±50.00% both upper and lower limit. |
Indication of comparator’s comparison result | LED indication LO/GO/HI and buzzer |
Control signal | Remote start input:”L” [0V] →”H” [DC12V] start |
Remote hold input:Open and “H” [DC12V] :Free run/”L” [0V]:Hold | |
comparison output [LO/GO/HI]:open collector:max.40V, 100mA | |
contact error output [CE]:open collector:max.40V, 100mA | |
end of comparison output [EOC]:open collector:max.40V, 100mA | |
Operation condition | [Temp.] +5℃~+40℃ [Humidity] less than 85% |
Power supply | AC100V/117V/220V/240V selectable, 50/60Hz, about 60VA |
Outer dimension | about 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.) |
Weight | about 3.8kg |
The Outline
AX-1136A is compared the resistance value of measuring thermistor with external standard thermistor and displayed tolerance by parcentage. And at the same time, the measred value is made the comparative judgement with the set value, and then is made the judged outpu.