Best Suitable for B, C, D, F, G, J, K classes, chip, melf, lead type resistance of sorting machine, and taping machine
● Ultra-precision:10ppm resolution, Basic accuracy ±0.005%±1digit
● High stability by the improvement of noise immunity for the isolated curcuit between analog part and digital part.
● Ultra-high stability by the improvement of noise immunity, isolated curcuit between analog part and digital part.
● Available to make the high speed and ultra-high stable measurement by the setting function of average time on measuring value for each range.
● Range of measurement for absolute value:0.00mΩ~125.00MΩ for %:10mΩ~100MΩ [±10.000%/-99.99%~+25.00%]
● Available to select the function for contact check before or after the measurement, or function of non-contact check.
● RS-232C and Centronics interface are built-in as standard equipment. [GP-IB is option]
● Printer output as standard equipment. (Centronics)
● Transfer function of Setting data is built-in as standard equipment. (Available to transfer the same setting data to another set of AX-163B)
● The checking circuit of the abnormal measuring current and voltage is built-in.
● Measuring probe self-cleaning circuit [utility model]
Specifications
Measuring range and Accuracy (at23℃±5℃), In a case of 180 days after calibration for half time, after 1 year for double
Measuring range | Setting range of standard value | Measuring current | Measuring accuracy | ||
---|---|---|---|---|---|
SLOW [Hi-Reso.] | SLOW [Lo-Reso.] | FAST [Lo-Reso.] | |||
100mΩ | 5mΩ~109mΩ | 180mA | within ±0.01%±α ±10d |
within ±0.02%±α ±2d |
within ±0.03%±2α±3d ±[2/(1+n)] d |
1Ω | 109.1mΩ~1.09Ω | 180mA | within ±0.005%±α ±3d |
within ±0.02%±α ±1d |
within ±0.02%±α±2d ±[2/(1+n)] d |
10Ω | 1.091Ω~10.9Ω | 90mA | |||
100Ω | 10.91Ω~109Ω | 18mA | within ±0.005% ±1d |
within ±0.02%±1d | within ±0.02%±2d ±[1/(1+n)] d |
1kΩ | 109.1Ω~1.09kΩ | 9mA | |||
10kΩ | 1.091kΩ~10.9kΩ | 0.9mA | |||
100kΩ | 10.91kΩ~109kΩ | 90μA | |||
1MΩ | 109.1kΩ~1.09MΩ | 9μA | within ±0.007% ±1d |
within ±0.05%±2d ±[1/(1+n)] d |
|
10MΩ | 1.091MΩ~10.9MΩ | 0.9μA | within ±0.02% ±10d |
within ±0.03%±1d | within ±0.2%±4d ±[1/(1+n)] d |
100MΩ | 10.91MΩ~109MΩ | 0.09μA | within ±0.1% ±20d |
within ±0.1%±2d | ---- |
d:digits n:Averaging time(msec), α=[100/setting value (mΩ)]×0.01%, In case of the measurement of absolute value, add ±1digit.
※The above FAST is accuracy under the condition shielded completely.
Measuring time | Remote start | Free running | |
---|---|---|---|
SLOW | FAST | SLOW | FAST |
AC1 period +about 7.5msec. ~about 408msec. |
about 11.5msec. ~about 408msec. |
about 5 times/sec. ~about 2 times/sec. |
about 10 times/sec. ~about 2 times/sec. |
EOC [End of comparison] pulse width | 1~250msec. or continuative |
---|---|
Measuring method | 2 or 4-terminal measurement |
Setting range value judgment | % Measurement:±10.000%/-99.99%~+25.00% Absolute measurements:00000~12500 |
Operation condition | [Temp.] 0℃~+50℃ [Humidity] less than 80% |
Power supply | AC85V~265V, 50~60Hz, about 50VA |
Outer dimension | about 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.) |
Weight | about 3.8kg |
The Outline
AX-163B can easily measure a wide range of resistance from 0.00mΩ to 125MΩ, high speed, and high accuracy.
The unit is installed microprocessor to judge the measured value for HI/GO/ LO, outputs signal outside.
The measured value is indicated as ±10.000%/-99.99%~+25.00%, or resistance value (max.12500 count).
The unit can be switched a measuring speed, both FAST and SLOW can set an integral time each range.
Contact-check function as a standard equipment: When measure 4 terminals, either one contact failure occurs, judges H decision, and outputs HI signal with C.E (Contact Error) signal outside regardless of its measured value at the same time.
The Contact-check is selectable from premeasurement/postmeasurement/OFF.
Besides, a check function to observe the measuring electrical current/electrical voltage always has set, when bad contact situation occurs during measurement, outputs to judge NG decision. Printed-out (based on centronics), RS-232C interface, and set date transfer functions are equipped as standard. Printed-out can be memorized 10,000 of the measuring date, the unit can be measured even though during printing out. Besides it also can be printed out a result of statistical analysis date for 100 unit of its measuring date.
RS-232C (GP-IB is an option) can be set measurement condition such as output the measuring date, standard resistance or limit value, and others.
The function of transfers the setup date can easily and quickly transfers another AX-163B for the same value of a range, standard, limit, is very useful when use two units together (joint cable is sold separately)
Besides, for setup of a measuring range to handle a key of front panel is only put in a standard resistance value to select automatically an appropriate range built in microprocessor so as to remove the burden of setup a range.
Moreover, a content of the setup keeps by battery backup even the power source OFF.