Ambient Zero Air Generator
Overview
AQMS-100
Zero Air Generator
FPI AQMS-100 provides ultra pure pollutant-free zero air for analytical purpose. Remote monitoring and configuration can also be achieved through digital access.
Key Features
- Ultra high purity zero air output
- Integrated dew point sensor
- Regenerative dryer utilizing molecular sieve to perform maintenance free advantage
- High performance scrubbers to remove hydrocarbon, CO, SO2, NO, NO2 and Ozone
- Remote display and configuration on key functions
Principle
The AQMS-100 generates clean and dry zero air by removing components which will cause interference on monitoring results.
Regenerative scrubber
The scrubber is filled with molecular sieve which has multiple holes and filters only water molecule with certain diameter. A second molecular sieve will be alternate for raw air diversion while the other is under regeneration by injecting compressed air to remove moisture.
HC/CO/NO scrubber
To remove hydrocarbon, CO and NO, three dependent scrubbers applying catalytic reaction are used, where HC and CO will be converted into CO2 which does not interfere analysis and NO will
be converted into NO2 which will then be removed by purifier.
Technical Data
Output | 20 SLPM at 30 psig |
Dew Point | -30℃ |
Output Concentration | SO2: <0.025 ppb |
NO: <0.025 ppb | |
NO2: <0.025ppb | |
O3: <0.3 ppb | |
CO: <10 ppb | |
CH4: <5 ppb | |
Other HC: <0.25ppb | |
Power Requirement | 100~240 VAC, Converter applicable |
Operating Temperature | 0~40℃ |
Dimensions | 221mm(H) x 482mm(W) x 554(D) |